232 ChromGate® Preparative Option<br />

These settings enable someone to consider the time delay for<br />

transporting the sample from the flow cell to the collector valve. If the<br />

option Use recycling valve is checked, the capillary from the flow cell via<br />

the recycling valve to the fractionation valve must be taken into account.<br />

If you activate the option Fixed time you can enter this time delay<br />

directly. This value will be valid, even if the flow rate is changed!<br />

Entering the Capillary Volume directly or calculating it by entering the<br />

Capillary Length and Inner Diameter, the delay time will be adapted to<br />

the programmed flow rate of the method.<br />

Solvent and/or Peak Recycling<br />

For an optional solvent and/or peak recycling, for most fraction collectors<br />

an additional recycling valve is necessary. As a recycling valve only a<br />

Knauer 6-port valve can be used. The Knauer fraction collector 3050<br />

includes a valve for recycling. For a multi valve fraction collector some<br />

ports of the fractionation valve can be defined for the recycling.<br />

To define a separate recycling valve, check the option Use recycling<br />

valve and click on the button. For the separate recycling valve a COM<br />

port or the IP address is to define. The settings for the volumetric delay<br />

will be displayed without access for changing. The settings are valid as<br />

well for the recycling valve as for the fractionation valve. Therefore, the<br />

capillary between the recycling valve and the fraction collector should be<br />

as short as possible.<br />

Only for the Knauer fraction collector 3050 and the multi valve fraction<br />

collectors the option Use own ports for recycling will be displayed. If<br />

you want to use this option, click into the check box for this option. For<br />

the multi valve fraction collector you will find, that the number of available<br />

ports for fractionation will be reduced by 2. Additionally the ports numbers<br />

for solvent and peak recycling will be shown. It is not possible only to<br />

enable one of the recycling options.<br />

Fig. 322 Recycling valve setup