Creating an Instrument Control Method 97<br />

Fig. 141 Detector K-2600 <strong>Advanced</strong> setup window<br />

Interpolate peak profile after scan<br />

While scanning, the data acquisition of the detector is interrupted. This<br />

leads to a plateau shape for that time and thus to a false quantitative<br />

result. When this option is activated the software interpolates a<br />

reasonable course of the chromatogram. The option is limited for runs<br />

with a time constant ≤ 1s.<br />

Warning! This option leads to a better shaped chromatogram;<br />

however the quantitative results are still not fully reliable. For<br />

accurate and reliable qualitative and quantitative results the run<br />

should be performed twice, with and without scanning.<br />

Scan if a peak is detected<br />

At any detected peak the software will perform a scan automatically if this<br />

option is activated. All further settings in this area are valid for the whole<br />

method. You may decide whether the scan shall be performed At the<br />

peak apex or At the beginning of the peak with the given Delay [min].<br />

With the Slope [µAU/sec]: and at Height [µAU]: settings you can define<br />

the thresholds, necessary for evaluating the signal as a peak. The unit<br />

[µAU] depends on the configuration setting.<br />

Use channel defines the data channel where the analysis is performed.<br />

You cannot change this selection during a run. However you can change<br />

the measuring wavelength of this channel as often as necessary.<br />

“Scan” column in time table<br />

If the option “Acquire 3D data” is enabled, a new column “Scan” will be<br />

available in the time table for the detector K-2600. If you want to perform<br />

a time-defined scan, enter a new line into the table with the desired time<br />

and enable the scan by clicking the check box. Please note, that during a<br />

scan the measurement signal(s) cannot be acquired. In the<br />

chromatogram you will find a plateau for the scanning time. It is not<br />

allowed to enable a scan in the first and last line of the time table. The<br />

time between two scan must be 0.33 min in minimum.<br />

Fig. 142 Detector K-2600 time based scan