Creating an Instrument Control Method 145<br />

Instrument Status – Detectors<br />

Instrument Status – RI Detectors (S 23[4]00, K-23[4]00/1)<br />

The tabs for all RI detectors appear identical, independent of the features<br />

of the detector. Unavailable features or functions are only grayed and<br />

inaccessible.<br />

Fig. 217 RI-Detector status tab, example: S 2300 / S 2400<br />

Monitor<br />

Run Status<br />

The status HOLD will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

Signal (µRIU)<br />

The actual signal is indicated in this field. The unit corresponds to the<br />

setting in the configuration.<br />

Direct Control<br />

The direct control option must be activated (see Fig. 249 on page 170).<br />

The buttons Flush and Autozero can be pressed to perform the<br />

corresponding functions. The flush function (flushing for 1 min) is<br />

deactivated during a run. The Autozero defines the current signal as a<br />

zero level.<br />

Instrument Status – UV Detectors<br />

(S 2520, S 2500, S 200, K-200, K-2000/1, K-2500/1)<br />

The tabs for most of the UV detectors appear identical, independent of<br />

the features of the detector. Unavailable features or functions are only<br />

grayed and inaccessible.