268 SEC Option<br />

Fig. 379<br />

Enable Analysis<br />

Click this box to enable SEC calculations for the current channel. Each<br />

channel can have its own analysis conditions.<br />

Molecular Mass<br />

For each calibration peak, enter a Molecular Mass that is representative<br />

of the calibration standard peak.<br />

Average Time (min)<br />

Enter the time in minutes of the calibration standard peak. If you have<br />

used the graphical Define SEC Peaks command, the peak times will be<br />

entered already for you. Note: the time is converted to seconds for<br />

calibration and Molecular Mass calculations.<br />

Time Slice<br />

Select a time slice width (in seconds) for the calculations. The slice width<br />

will determine the frequency of Molecular Mass fractions reported. If a<br />

Slice Report is generated, it will be based on the slice width you select<br />

here. The software will force this number to be an integral of the data<br />

collection rate during the analysis.<br />

Baseline Start and Stop Times<br />

The baseline start and stop times (in minutes) can be entered manually<br />

or selected graphically using the SEC buttons on the chromatogram<br />

toolbar.<br />

Results Range Start and Stop<br />

One or more ranges of the chromatogram must be defined to be used in<br />

calculation and reporting of the Molecular Mass distribution. Each range<br />

is determined by the results range start and stop values. The times (in<br />

minutes) can be entered manually or graphically using a stored<br />

chromatogram.