126 Creating an Instrument Control Method<br />

starts. The programmed volume of wash solvent is used to wash the<br />

needle at the wash position.<br />

The wash position may be contaminated with the contents of the<br />

buffer tubing, which may generate cross-contamination. To<br />

prevent contamination of the wash position, program a<br />

“dispense to waste” action before programming a wash action.<br />

Valve<br />

The positions of the installed high pressure valves are programmed in<br />

this screen. The injector valve has two positions: INJECT and LOAD.<br />

If the ISS option is installed, the two valves have the two positions: 1-6<br />

and 1-2.<br />

Wait<br />

The WAIT action can be used to let the autosampler wait for a<br />

programmed time with maximum of 9 h 59 min 59 sec.<br />

During WAIT the needle turns back to home position if the next<br />

action is not at the same position. Should the needle be required<br />

to remain in the current position, it is necessary to program a<br />

dummy step after the WAIT in which an aspirate or dispense<br />

action of 0 µl is programmed at the actual position.<br />

Compressor<br />

With the compressor action the compressor is activated to transfer air<br />

pressure to a vial. The compressor will stay active until it is switched<br />

OFF. The compressor will be automatically switched off at the end of<br />

the needle wash routine if a needle wash is used.<br />

Auxiliaries<br />

With the Auxiliaries action the four standard auxiliaries (contact<br />

closures) can be controlled (see time base events).<br />

Wait for Input<br />

The autosampler will wait for one of the four inputs to become HIGH<br />

or LOW before continuing with the next step.<br />

Prog. Output<br />

The Prog. Output action is used to define two programmable outputs<br />

(contact closures). These are similar to the auxiliaries.<br />

Code Output<br />

The Code Output action is used to program the output to the<br />

connector P3 TIMED OUTPUTS on the rear panel of the autosampler.<br />

This is a HEX output in the range 0 to 15 (F).<br />

Markers<br />

The markers normally generated by the autosampler are not active in<br />

the user defined method, but can be programmed in this screen. The<br />

choices are: Inject, Vial, and Labeled.<br />

Markers must be programmed, because they are not automatically<br />

generated as in the regular injection method. Refer to the<br />

Triathlon/Endurance User <strong>Manual</strong> for the cable connections.<br />

SSV Port<br />

Program the solvent selection valve (SSV) port position, range 1 to 6.<br />

Repeat<br />

The last x steps will be repeated y times.<br />

Button functions:<br />

Click the Add button to complete the operation.