126 Creating an Instrument Control Method<br />
starts. The programmed volume of wash solvent is used to wash the<br />
needle at the wash position.<br />
The wash position may be contaminated with the contents of the<br />
buffer tubing, which may generate cross-contamination. To<br />
prevent contamination of the wash position, program a<br />
“dispense to waste” action before programming a wash action.<br />
Valve<br />
The positions of the installed high pressure valves are programmed in<br />
this screen. The injector valve has two positions: INJECT and LOAD.<br />
If the ISS option is installed, the two valves have the two positions: 1-6<br />
and 1-2.<br />
Wait<br />
The WAIT action can be used to let the autosampler wait for a<br />
programmed time with maximum of 9 h 59 min 59 sec.<br />
During WAIT the needle turns back to home position if the next<br />
action is not at the same position. Should the needle be required<br />
to remain in the current position, it is necessary to program a<br />
dummy step after the WAIT in which an aspirate or dispense<br />
action of 0 µl is programmed at the actual position.<br />
Compressor<br />
With the compressor action the compressor is activated to transfer air<br />
pressure to a vial. The compressor will stay active until it is switched<br />
OFF. The compressor will be automatically switched off at the end of<br />
the needle wash routine if a needle wash is used.<br />
Auxiliaries<br />
With the Auxiliaries action the four standard auxiliaries (contact<br />
closures) can be controlled (see time base events).<br />
Wait for Input<br />
The autosampler will wait for one of the four inputs to become HIGH<br />
or LOW before continuing with the next step.<br />
Prog. Output<br />
The Prog. Output action is used to define two programmable outputs<br />
(contact closures). These are similar to the auxiliaries.<br />
Code Output<br />
The Code Output action is used to program the output to the<br />
connector P3 TIMED OUTPUTS on the rear panel of the autosampler.<br />
This is a HEX output in the range 0 to 15 (F).<br />
Markers<br />
The markers normally generated by the autosampler are not active in<br />
the user defined method, but can be programmed in this screen. The<br />
choices are: Inject, Vial, and Labeled.<br />
Markers must be programmed, because they are not automatically<br />
generated as in the regular injection method. Refer to the<br />
Triathlon/Endurance User <strong>Manual</strong> for the cable connections.<br />
SSV Port<br />
Program the solvent selection valve (SSV) port position, range 1 to 6.<br />
Repeat<br />
The last x steps will be repeated y times.<br />
Button functions:<br />
Click the Add button to complete the operation.