12.02.2013 Views

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

ChromeGate 3.3.2 Software Manual - KNAUER Advanced Scientific ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

102 Creating an Instrument Control Method<br />

Fig. 150 DAD setup, scan range<br />

This section allows you to specify the scan spectral interval, in nm. The<br />

given range limits are read out from the used instrument. Therefore they<br />

may differ from the shown values.<br />

Start<br />

Enter the start wavelength for the scan, in nm.<br />

End<br />

Enter the end wavelength for the scan, in nm.<br />

Bandwidth<br />

The Bandwidth is the wave length range that will be used to measure a<br />

data point. For the spectral acquisition it is highly recommended to use a<br />

bandwidth of 1 nm.<br />

The scan section limits Start and End are important even if no PDA<br />

option is available. All of the measuring wavelengths in the<br />

Wavelength Program for the used channels 1 to 4 must be within<br />

these limits.<br />

Signal Mode<br />

The active signal mode is already selected in the instrument configuration<br />

(see page 50). The choices are Absorption and Intensity. For the<br />

Kontron DAD 540 it is fixed on Absorption.<br />

Acquire 3D data<br />

You have access to this selection field only with the PDA option.<br />

Wavelength Program<br />

Fig. 151 Detector S 2600 wavelength program table<br />

Used Channel<br />

Check the box(es) to turn the acquisition for the selected channel(s) on. If<br />

the box is not selected, no data will be acquired and no time program will<br />

be executed for this channel. As many channels are activated (at least<br />

one) as many wavelength columns appear in the spreadsheet below.<br />

Bandwidth [nm]<br />

The bandwidth is a measure of the range over which the calculations and<br />

measurements are used to calculate the intensity for a particular<br />

wavelength. For example, with a 10 nm bandwidth, the intensity for 250<br />

nm would be calculated from 245 nm to 255 nm.<br />

A wider bandwidth results in an increased signal to noise ratio because<br />

the multiple measurements cause the random noise to be cancelled out.<br />

That is, if the noise is truly random, more measurements will decrease

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!