148 Creating an Instrument Control Method<br />

select a time constant, which is at least one tenth of the peak width at the<br />

base of the narrowest peak of interest.<br />

The range can be set to define the range of the analog detector outputs.<br />

The choice is 0.100, 0.200, 0.500, 1.000, and 2.000 Absorption Units Full<br />

Scale (AUFS).<br />

The Kontron x30 and x35 detectors can be used in a dual channel mode.<br />

As far as the option Used Channel #2 is checked, the otherwise not<br />

accessible cycle time becomes active for changing. The cycle time you<br />

can set within the range of 0.40 through 1.00 seconds with 0.05 s<br />

increments.<br />

For the x32 detectors the option Used Channel #2 is not available and<br />

the cycle time is not displayed at all.<br />

Instrument Status – fast scanning UV Detector (K-2600)<br />

Fig. 222 Detector status tab, K-2600<br />

Monitor<br />

Run Status<br />

The status HOLD will be displayed when no program is running;<br />

otherwise the Run Status displays a run time in minutes elapsed from the<br />

start of the run.<br />

Digital Outs (Events)<br />

The current state for each digital output will be displayed in this control.<br />

The On state is represented by an active red Light Emission Diode LED<br />

symbol ( ), the Off state, including waiting for pulse, is represented by<br />

an inactive grey LED symbol ( ).<br />

Channel 1 / 2 / 3 / 4<br />

The actual signal in the units which have been chosen in the detector<br />

configuration dialog at the actual wavelength in nm is displayed for each<br />

active channel.<br />

GLP<br />

The GLP data of the detector are displayed in this area. The data will be<br />

actualized by pressing the Refresh button.<br />

Direct Control<br />

The direct control option must be activated (see Fig. 249 on page 170).<br />

Lamp<br />

The On state of the lamp is represented by an active green LED symbol<br />

( ), the Off state is represented by an inactive gray LED symbol ( ).<br />

Clicking on the Lamp Off button will (after inquiry) switch the lamp off<br />

and change the button label to Lamp ON.