22.05.2022 Views

DƯỢC LÍ Goodman & Gilman's The Pharmacological Basis of Therapeutics 12th, 2010

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

and left-to-right shunting of cardiac output, oxygen supplementation

must be regulated carefully because of the risk of further reducing

pulmonary vascular resistance and increasing pulmonary blood flow.

Metabolism. Inhalation of 100% O 2

does not produce detectable

changes in O 2

consumption, CO 2

production, respiratory quotient,

or glucose utilization.

Oxygen Administration

Oxygen is supplied as a compressed gas in steel cylinders,

and a purity of 99% is referred to as medical

grade. Most hospitals have oxygen piped from insulated

liquid oxygen containers to areas of frequent use.

For safety, oxygen cylinders and piping are color-coded

(green in the U.S.), and some form of mechanical

indexing of valve connections is used to prevent the

connection of other gases to oxygen systems. Oxygen

concentrators, which employ molecular sieve, membrane,

or electrochemical technologies, are available for

low-flow home use. Such systems produce 30-95%

oxygen depending on the flow rate.

Oxygen is delivered by inhalation except during

extracorporeal circulation, when it is dissolved directly

into the circulating blood. Only a closed delivery system

with an airtight seal to the patient’s airway and complete

separation of inspired from expired gases can precisely

control FI O2

. In all other systems, the actual delivered

FI O2

will depend on the ventilatory pattern (i.e., rate, tidal

volume, inspiratory–expiratory time ratio, and inspiratory

flow) and delivery system characteristics.

Low-Flow Systems. Low-flow systems, in which the oxygen flow is

lower than the inspiratory flow rate, have a limited ability to raise the

FI O

because they depend on entrained room air to make up the balance

of the inspired gas. The FI O

2

of these systems is extremely sensitive

to small changes in the ventilatory pattern. Nasal

2

cannulae—small, flexible prongs that sit just inside each naris—

deliver oxygen at 1-6 L/min. The nasopharynx acts as a reservoir for

storing the oxygen, and patients may breathe through either the

mouth or nose as long as the nasal passages remain patent. These

devices typically deliver 24-28% FI at 2-3 L/min. Up to 40% O 2 FIO 2

is possible at higher flow rates, although this is poorly tolerated for

more than brief periods because of mucosal drying. The simple face

mask, a clear plastic mask with side holes for clearance of expiratory

gas and inspiratory air entrainment, is used when higher concentrations

of oxygen delivered without tight control are desired. The

maximum FI O

of a face mask can be increased from around 60% at

2

6-15 L/min to > 85% by adding a 600- to 1000-mL reservoir bag. With

this partial rebreathing mask, most of the inspired volume is drawn

from the reservoir, avoiding dilution by entrainment of room air.

High-Flow Systems. The most commonly used high-flow oxygen

delivery device is the Venturi-style mask, which uses a specially

designed mask insert to entrain room air reliably in a fixed ratio and

thus provides a relatively constant FI O

at relatively high flow rates.

2

Typically, each insert is designed to operate at a specific oxygen flow

rate, and different inserts are required to change the FI . Lower

O 2

delivered FI O

values use greater entrainment ratios, resulting in

2

higher total (oxygen plus entrained air) flows to the patient, ranging

from 80 L/min for 24% FI to 40 L/min at 50% O

. While these

2 FIO 2

flow rates are much higher than those obtained with low-flow

devices, they still may be lower than the peak inspiratory flows for

patients in respiratory distress, and thus the actual delivered O 2

concentration

may be lower than the nominal value. Oxygen nebulizers,

another type of Venturi-style device, provide patients with

humidified oxygen at 35-100% FI O

at high flow rates. Finally, oxygen

blenders provide high inspired oxygen concentrations at very

2

high flow rates. These devices mix high-pressure compressed air and

oxygen to achieve any concentration of O 2

from 21-100% at flow

rates of up to 100 L/min. These same blenders are used to provide

control of FI O

for ventilators, CPAP/BiPAP machines, oxygenators,

2

and other devices with similar requirements. Again, despite the high

flows, the delivery of high FI O

to an individual patient also depends

2

on maintaining a tight-fitting seal to the airway and/or the use of

reservoirs to minimize entrainment of diluting room air.

Monitoring of Oxygenation. Monitoring and titration are required

to meet the therapeutic goal of oxygen therapy and to avoid complications

and side effects. Although cyanosis is a physical finding of

substantial clinical importance, it is not an early, sensitive, or reliable

index of oxygenation. Cyanosis appears when ~5 g/dL of deoxyhemoglobin

is present in arterial blood, representing an oxygen saturation

of ~67% when a normal amount of hemoglobin (15 g/dL) is

present. However, when anemia lowers the hemoglobin to 10 g/dL,

then cyanosis does not appear until the arterial blood saturation has

decreased to 50%. Invasive approaches for monitoring oxygenation

include intermittent laboratory analysis of arterial or mixed venous

blood gases and placement of intravascular cannulae capable of continuous

measurement of oxygen tension. The latter method, which

relies on fiber-optic oximetry, is used frequently for the continuous

measurement of mixed venous hemoglobin saturation as an index of

tissue extraction of oxygen, usually in critically ill patients.

Noninvasive monitoring of arterial oxygen saturation can be

achieved using transcutaneous pulse oximetry, in which oxygen saturation

is measured from the differential absorption of light by oxyhemoglobin

and deoxyhemoglobin and the arterial saturation

determined from the pulsatile component of this signal. Application

is simple, and calibration is not required. Pulse oximetry measures

hemoglobin saturation and not PO 2

. It is not sensitive to increases in

PO 2

that exceed levels required to saturate the blood fully. Pulse

oximetry is very useful for monitoring the adequacy of oxygenation

during procedures requiring sedation or anesthesia, rapid evaluation

and monitoring of potentially compromised patients, and titrating

oxygen therapy in situations where toxicity from oxygen or side

effects of excess oxygen are of concern. Near infrared spectroscopy

(NIRS) is a noninvasive technique being used to monitor oxygen

content in the cerebral cortex. Unlike pulse oximetry NIRS measures

all reflected light in both pulsatile arterial blood and nonpulsatile

venous blood, the primary compartment in the cerebral

vascular bed. NIRS is useful to monitor cerebral oxygenation in surgical

procedures involving cardiopulmonary bypass and circulatory

arrest (Guarracino, 2008).

Complications of Oxygen Therapy. Administration of supplemental

oxygen is not without potential complications. In addition to the

555

CHAPTER 19

GENERAL ANESTHETICS AND THERAPEUTIC GASES

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!