JSN M7860<br />

Name Monitoring System, Cardiac Catheterization Lab<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6515<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 20<br />

Amps2 0<br />

Weight (lbs) 375 Weight (kgs) 171<br />

Height (in) 50 Height (cen) 127<br />

Width (in) 65 Width (cen) 165<br />

Depth (in) 31 Depth (cen) 79<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 3623<br />

Watts1 2400<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Miscellaneous<br />

Computerized monitoring system for use in cardiac catheterization labs and cardio-thoracic operating rooms. <strong>The</strong> system can display up<br />

to 32 different waveforms simultaneously, perform calculations of hemodynamic parameters, valve areas and cardiac output, archive<br />

procedure results and accept inputs from a variety of monitoring devices. <strong>The</strong> installed system can also include remote<br />

monitors/terminals and additional workstations and may interface with hospital-wide physiological monitoring or clinical information<br />

systems. <strong>The</strong> system includes several configurations in different locations. Database physical dimensions reflect the largest of these work<br />

centers. <strong>The</strong> estimated total 20 amp electrical requirements will be spread across several circuits in several locations. <strong>The</strong> system price<br />

varies greatly on the size and sophistication of the end user's requirements.<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />