JSN S0237<br />

Name Sterilizer,ELEC,VAC,2DO,RCSD 1WLL, 20x20x38 Cham<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army A | Contractor Furnished, Contractor Installed<br />

Navy A | Contractor Furnished, Contractor Installed<br />

USAF A | Contractor Furnished, Contractor Installed<br />

VA A | Contractor Furnished, Contractor Installed<br />

IHS A | Contractor Furnished, Contractor Installed<br />

Utilities<br />

NSN 6530<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing D | Cold and hot water and drain<br />

Volts1 120<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 3<br />

Amps1 12<br />

Amps2 84<br />

Weight (lbs) 1760 Weight (kgs) 799<br />

Height (in) 75 Height (cen) 190<br />

Width (in) 34 Width (cen) 86<br />

Depth (in) 46 Depth (cen) 117<br />

Volts2 208<br />

Hertz 50/60<br />

Hertz Dep 0<br />

Hertz Switch -1<br />

BTU / Hour 9550<br />

Watts1 1440<br />

Watts2 30262<br />

Description<br />

Unit of Issue EA<br />

Group Sterilizers and Associated <strong>Equipment</strong><br />

A recessed mounted (through one wall), double power doors (vertical sliding), vacuum, sterilizer with integral steam generator. <strong>The</strong> unit is<br />

controlled by a microcomputer that monitors and controls all unit operations and functions and provides both audible and visual<br />

indications of deviations. A printer-recorder documents and records each cycle performance. For general purpose pre-vacuum or gravity<br />

steam sterilization of hospital and laboratory supplies at temperatures in the range from 110°C to 135°C (230°F to 275°F). A liquid cycle is<br />

also provided at temperatures in the range from 100°C to 121°C (212°F to 250°F). NOTE: Various electrical configurations are available for<br />

the integral steam generator, for pricing purposes 208V, 50/60HZ, 3PH, 84AMPS is used.<br />