JSN L7070<br />

Name Spectrophotometer, Atomic Absorption<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6640<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas D | Air, high pressure (85 PSIG) max<br />

Utility4: Misc Gas C | Nitrous oxide<br />

Utility5: Non-Med Gas H | Acetylene gas<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 3<br />

Amps2 0<br />

Weight (lbs) 216 Weight (kgs) 98<br />

Height (in) 26 Height (cen) 65<br />

Width (in) 54 Width (cen) 137<br />

Depth (in) 26 Depth (cen) 65<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch -1<br />

BTU / Hour 0<br />

Watts1 360<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Laboratory <strong>Equipment</strong><br />

Atomic absorption spectrophotometer. This is a double beam unit with a graphite furnace for background absorption. It has an<br />

automatic lamp selection, corrosion resistant materials construction, high resolution graphics, wavelength ranges from 190 to 900nm, a<br />

flame atomizer with pneumatic nebulization, the ability to accommodate an impact bead or a flow spoiler and an automatic calibration<br />

capability. Used in laboratories to process samples for analysis.<br />