05.04.2013 Views

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

JSN X6175<br />

Name Radiographic/Fluoro Unit, Angio, Biplane, Digital<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army F | Government Furnished, Government Installed<br />

Navy F | Government Furnished, Government Installed<br />

USAF F | Government Furnished, Government Installed<br />

VA F | Government Furnished, Government Installed<br />

IHS F | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6525<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 480<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc B | Lead lined walls<br />

Phase1 3<br />

Phase2 0<br />

Amps1 27<br />

Amps2 0<br />

Weight (lbs) 17455 Weight (kgs) 7925<br />

Height (in) 0 Height (cen) 0<br />

Width (in) 0 Width (cen) 0<br />

Depth (in) 0 Depth (cen) 0<br />

Volts2 0<br />

Hertz 50/60<br />

Hertz Dep 0<br />

Hertz Switch -1<br />

BTU / Hour 67135<br />

Watts1 22447<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group X-Ray <strong>Equipment</strong> and Accessories<br />

This system is specifically designed to perform biplane radiographic/fluoroscopic examinations in the Special Procedures Department. Online<br />

digital angio image processing will provide instant availability of images for review. This units characteristics and components include:<br />

100 kW micro-processor controlled X-ray generators, C-arm and U arm with 9” multi-field Image Intensifier, integrated X-ray tube unit.<br />

<strong>The</strong> Digital Spot Imaging for both the AP and Lateral planes shall consist of a computer, keyboard with acquisition and viewing monitor<br />

and a slave monitor. <strong>The</strong> system shall be DICOM 3.0 compatible, for easy linkage to filmless image management systems and review<br />

stations. It is recommended that the TV monitors be ceiling suspended.<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />

Page 1381 of 1462