JSN M8610<br />

Name Endoscopy Cart, Video (CCD) Scope<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6515<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical A | 120 volt, conventional outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas E | Carbon dioxide gas<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 21<br />

Amps2 0<br />

Weight (lbs) 200 Weight (kgs) 91<br />

Height (in) 72 Height (cen) 183<br />

Width (in) 36 Width (cen) 91<br />

Depth (in) 36 Depth (cen) 91<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 2520<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Miscellaneous<br />

Video support system for endoscopes which have a charge coupled device (CCD) camera chip or similar technology embedded in the<br />

distal end of the endoscope. This system does not support direct observation fiberoptic endoscopes. Refer to JSNs M8505-M8507 for<br />

fiberoptic endoscope carts. A typical cart system includes the cart, a light source, an insufflator, a suction unit, a heat probe unit, an<br />

electrosurgical apparatus, a video controller, a monitor, a video recorder and a color printer. This JSN does not include the endoscope;<br />

refer to the endoscopes at JSNs M8500-M8550. Each cart can support one or more types of video endoscope and should be specifically<br />

tailored to its intended use(s). This cart can be configured to interface with a network endoscopy information management system; refer<br />

to JSN M8600. Database physical information and pricing is for a higher cost system containing one of each of the above components.<br />

Electrical draw may require two separate electrical circuits to support one cart.<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />