JSN L1770<br />

Name Ultracentrifuge<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6640<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 220<br />

Utility2: Electrical C | 208/220 volt<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 20<br />

Amps2 0<br />

Weight (lbs) 1025 Weight (kgs) 466<br />

Height (in) 48 Height (cen) 122<br />

Width (in) 37 Width (cen) 94<br />

Depth (in) 28 Depth (cen) 71<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 3400<br />

Watts1 4400<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Laboratory <strong>Equipment</strong><br />

Ultracentrifuge. <strong>The</strong> unit handles tubes from 0.2ml to 3.5ml, has a programmable memory for repeat of select runs and has selectable<br />

acceleration and deceleration rates. It operates in the range from 2-40 degrees C. <strong>The</strong> unit has a digital timer, and electronic door locking<br />

mechanism. <strong>The</strong> maximum speed is 90,000 RPM. Applications range from use as a simple pelleting device to the quantitation of<br />

molecular species by sedimentation equilibrium measurements.<br />