JSN M4820<br />

Name Hemodialysis Unit, Mbl, w/Reverse Osmosis System<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6515<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical A | 120 volt, conventional outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 14<br />

Amps2 0<br />

Weight (lbs) 175 Weight (kgs) 80<br />

Height (in) 52 Height (cen) 132<br />

Width (in) 23 Width (cen) 58<br />

Depth (in) 26 Depth (cen) 66<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 1680<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Miscellaneous<br />

Mobile renal dialysis system with a reverse osmosis water purification unit. <strong>The</strong> system consists of three major components: the dialysate<br />

delivery system, the extracorporeal blood delivery circuit (blood pump) and the dialyzer. <strong>The</strong> system also has its own monitoring and<br />

control circuits with temperature controls. An alarm system is automatically activated when there is changes in pressure or in the<br />

existence of leaks. Multiple electrical circuits (one for the dialyzer, one for the reverse osmosis system) may be needed to power this<br />

equipment combination. This dialyzer is intended for self-contained use in intensive care or step down units which do not have the water<br />

purification machinery found in a hemodialysis clinic. See JSN M1715 for units used in a hemodialysis clinic.<br />