JSN M4865<br />

Name Reprocessor, Renal Dialyzer Cartridge<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 4100<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing E | Treated water and drain<br />

Volts1 120<br />

Utility2: Electrical A | 120 volt, conventional outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc E | Vent to atmosphere<br />

Phase1 1<br />

Phase2 0<br />

Amps1 15<br />

Amps2 0<br />

Weight (lbs) 127 Weight (kgs) 58<br />

Height (in) 16 Height (cen) 41<br />

Width (in) 15 Width (cen) 38<br />

Depth (in) 13 Depth (cen) 33<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 1800<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Miscellaneous<br />

Renal dialyzer disinfection system. Unit is used to flush and disinfect dialyzer cartridges after a hemodialysis session. <strong>The</strong> unit then refills<br />

the cartridge with preservative for storage until reuse by the same patient during a later session. <strong>The</strong> manufacturer recommends a<br />

dedicated 20 amp ground fault interrupt protected power circuit for this equipment. See manufacturer's pre-installation checklist for<br />

ventilation requirements. Use JSN M4850 to remove disinfectant from dialyzer cartridge prior to use.<br />