JSN X6185<br />

Name Rad/Fluoro Unit, Remote, 80 kW, 90/90 Table<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army F | Government Furnished, Government Installed<br />

Navy F | Government Furnished, Government Installed<br />

USAF F | Government Furnished, Government Installed<br />

VA F | Government Furnished, Government Installed<br />

IHS F | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6525<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 480<br />

Utility2: Electrical E | 440 volt, 3 phase<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc B | Lead lined walls<br />

Phase1 3<br />

Phase2 0<br />

Amps1 145<br />

Amps2 0<br />

Weight (lbs) 4967 Weight (kgs) 2255<br />

Height (in) 0 Height (cen) 0<br />

Width (in) 0 Width (cen) 0<br />

Depth (in) 0 Depth (cen) 0<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 12542<br />

Watts1 120548<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group X-Ray <strong>Equipment</strong> and Accessories<br />

This system is specifically designed to perform radiographic/fluoroscopic examinations in the Radiology Department. On-line digital<br />

image processing will provide instant availability of images for review via infrared remote control. This units characteristics and<br />

components include, 80 kW micro-processor controlled X-ray generator, remote-controlled 90/90 table with 15” multi-field Image<br />

Intensifier and integrated tube support. <strong>The</strong> Digital Spot Imaging consists of a computer, keyboard with acquisition, viewing monitor and<br />

a slave monitor. <strong>The</strong> system shall be DICOM 3.0 compatible, for easy linkage to filmless image management systems and review stations.<br />

It is recommended that the TV monitors be ceiling suspended.<br />