JSN S0125<br />

Name Sterilizer, Stm, VAC, 1DO, CAB, 16x16x26 Chamber<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army A | Contractor Furnished, Contractor Installed<br />

Navy A | Contractor Furnished, Contractor Installed<br />

USAF A | Contractor Furnished, Contractor Installed<br />

VA A | Contractor Furnished, Contractor Installed<br />

IHS A | Contractor Furnished, Contractor Installed<br />

Utilities<br />

NSN 6530<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing B | Cold water and drain<br />

Volts1 120<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas A | Steam<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 10<br />

Amps2 0<br />

Weight (lbs) 1100 Weight (kgs) 500<br />

Height (in) 75 Height (cen) 190<br />

Width (in) 40 Width (cen) 102<br />

Depth (in) 59 Depth (cen) 150<br />

Volts2 0<br />

Hertz 50/60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 4300<br />

Watts1 1200<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Sterilizers and Associated <strong>Equipment</strong><br />

A cabinet mounted, single power door (vertical sliding), vacuum, sterilizer utilizing facility supplied steam. <strong>The</strong> unit is controlled by a<br />

microcomputer that monitors and controls all unit operations and functions and provides both audible and visual indications of<br />

deviations. A printer-recorder documents and records each cycle performance. For general purpose pre-vacuum or gravity steam<br />

sterilization of hospital and laboratory supplies at temperatures in the range from 110°C to 135°C (230°F to 275°F). A liquid cycle is also<br />

provided at temperatures in the range from 100°C to 121°C (212°F to 250°F).<br />