JSN T0015<br />

Name Defibrillator Analyzer, Automatic, Portable<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6625<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical A | 120 volt, conventional outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 2<br />

Amps2 0<br />

Weight (lbs) 7 Weight (kgs) 4<br />

Height (in) 5 Height (cen) 13<br />

Width (in) 13 Width (cen) 33<br />

Depth (in) 11 Depth (cen) 28<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 240<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Shop <strong>Equipment</strong>/Tools<br />

Defibrillator analyzer. Characteristics/components include automatic and portable with direct reading; accurately verifies the output<br />

characteristics of all defibrillators while simulating the resistance of the human body; output energy measured in joules or watt seconds on<br />

two scales of 0-200.0 or 0-1000; output energy remains displayed until the reset button is depressed once the defibrillator is fired; and<br />

output jacks provided for oscilloscope connection to observe defibrillator pulse wave shape. Shall include the optional pacemaker testing<br />

module if required by the end user. Battery powered models are available.<br />