JSN M0765<br />

Name Regulator, Vacuum<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6515<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 0<br />

Utility2: Electrical . |<br />

Utility3: Med Gas B | Vacuum<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 0<br />

Phase2 0<br />

Amps1 0<br />

Amps2 0<br />

Weight (lbs) 1 Weight (kgs) 1<br />

Height (in) 6 Height (cen) 15<br />

Width (in) 3 Width (cen) 8<br />

Depth (in) 4 Depth (cen) 10<br />

Volts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Miscellaneous<br />

Vacuum pressure regulator for connection to central piped vacuum system. Standard display scale is graduated at least from 0 to 200 mm<br />

Hg of vacuum. Displays on specialized regulators may cover other vacuum ranges. Regulator type (continuous, intermittent,<br />

continuous/intermittent, surgical, pediatric, thoracic, etc.) as required. To be used in delivery, neonatal, pediatrics or any area where<br />

suction is required. Database pricing reflects continuous regulators graduated to 200 mm Hg with a full line vacuum selection switch and<br />

DISS configured inlets and outlets.<br />

Hertz<br />

Hertz Dep 0<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 0<br />

Watts2 0<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />