JSN X6340<br />

Name Rad/Fluoro Unit, 65 kW, 12" I.I., 90/15 Table<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army F | Government Furnished, Government Installed<br />

Navy F | Government Furnished, Government Installed<br />

USAF F | Government Furnished, Government Installed<br />

VA F | Government Furnished, Government Installed<br />

IHS F | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6525<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 480<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc B | Lead lined walls<br />

Phase1 3<br />

Phase2 3<br />

Amps1 117<br />

Amps2 140<br />

Weight (lbs) 6704 Weight (kgs) 3044<br />

Height (in) 0 Height (cen) 0<br />

Width (in) 0 Width (cen) 0<br />

Depth (in) 0 Depth (cen) 0<br />

Volts2 400<br />

Hertz 60<br />

Hertz Dep 0<br />

Hertz Switch -1<br />

BTU / Hour 12564<br />

Watts1 97270<br />

Watts2 96992<br />

Description<br />

Unit of Issue EA<br />

Group X-Ray <strong>Equipment</strong> and Accessories<br />

This system is specifically designed to perform radiographic/fluoroscopic examinations in the Radiology Department. <strong>The</strong> characteristics<br />

and components include a ceiling suspended 0.6/1.0 mm tube unit, 65 kW micro-processor controlled X-ray generator, 90/15 table with<br />

12” multi-field Image Intensifier (I.I.) and bucky wall stand. <strong>The</strong> digital system will be DICOM 3.0 or latest version compatible, for easy<br />