JSN S5500<br />

Name Sterilizer, Hydrogen Peroxide Gas, 1.75 cu ft Cham<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6530<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing . |<br />

Volts1 120<br />

Utility2: Electrical B | 120 volt, special outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 0<br />

Amps1 16<br />

Weight (lbs) 375 Weight (kgs) 171<br />

Height (in) 53 Height (cen) 135<br />

Width (in) 22 Width (cen) 56<br />

Depth (in) 33 Depth (cen) 84<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

Description<br />

Unit of Issue EA<br />

Group Sterilizers and Associated <strong>Equipment</strong><br />

A compact low temperature hydrogen peroxide gas plasma sterilizer mounted on a mobile cart. Sterilization chamber shall have a<br />

capacity of approximately 1.75 cubic feet. <strong>The</strong> sterilizer shall be capable of processing metal instruments, plastics, heat and moisture<br />

sensitive devices in approximately 45 minutes with no aeration required. Unit shall be equipped with a visual display and audible alarm. A<br />

built in printer will provide system performance record for each cycle.<br />

Amps2<br />

BTU / Hour<br />

Watts1 1920<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />

Page 1252 of 1462<br />

Watts2