JSN S0332<br />

Name Sterilizer,Stm,VAC,1DO,RCSD 1WLL,24x36x36 Chamber<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army A | Contractor Furnished, Contractor Installed<br />

Navy A | Contractor Furnished, Contractor Installed<br />

USAF A | Contractor Furnished, Contractor Installed<br />

VA A | Contractor Furnished, Contractor Installed<br />

IHS A | Contractor Furnished, Contractor Installed<br />

Utilities<br />

NSN 6530<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing B | Cold water and drain<br />

Volts1 120<br />

Utility2: Electrical F | Special electrical requirements<br />

Utility3: Med Gas D | Air, high pressure (85 PSIG) max<br />

Utility4: Misc Gas A | Steam<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc . |<br />

Phase1 1<br />

Phase2 1<br />

Amps1 5<br />

Amps2 15<br />

Weight (lbs) 3440 Weight (kgs) 1562<br />

Height (in) 76 Height (cen) 193<br />

Width (in) 59 Width (cen) 150<br />

Depth (in) 73 Depth (cen) 185<br />

Volts2 120<br />

Hertz 50/60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 12500<br />

Watts1 600<br />

Watts2 1800<br />

Description<br />

Unit of Issue EA<br />

Group Sterilizers and Associated <strong>Equipment</strong><br />

A recessed mounted (through one wall), single door (right or left-hinged, appropriate door configuration to be indicated on equipment<br />

elevation drawings), vacuum, sterilizer utilizing a remote electric steam generator. <strong>The</strong> unit is controlled by a microcomputer that monitors<br />

and controls all unit operations and functions and provides both audible and visual indications of deviations. A printer-recorder<br />

documents and records each cycle performance. For general purpose pre-vacuum or gravity steam sterilization of hospital and laboratory<br />

supplies at temperatures in the range from 110°C to 135°C (230°F to 275°F). A liquid cycle is also provided at temperatures in the range<br />

from 100°C to 121°C (212°F to 250°F). NOTE: For the electric steam generator, refer to JSN: S1830. Door may be power operated.<br />