JSN L1000<br />

Name Analyzer, Blood Chem, Auto, 60 Samples/Hour<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army C | Government Furnished, Government Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6640<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing E | Treated water and drain<br />

Volts1 120<br />

Utility2: Electrical B | 120 volt, special outlet<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc A | Earth ground<br />

Phase1 1<br />

Phase2 0<br />

Amps1 5<br />

Amps2 0<br />

Weight (lbs) 90 Weight (kgs) 41<br />

Height (in) 21 Height (cen) 53<br />

Width (in) 20 Width (cen) 51<br />

Depth (in) 21 Depth (cen) 53<br />

Volts2 0<br />

Hertz 50/60<br />

Hertz Dep 0<br />

Hertz Switch -1<br />

BTU / Hour 0<br />

Watts1 600<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Laboratory <strong>Equipment</strong><br />

Automatic blood/chemistry analyzer. It is a four parameter electrolyte analyzer. <strong>The</strong> unit measures Na, K, Cl concentrations using an ion<br />

selective electrode and CO2 concentrations using differential pH. <strong>The</strong> analyzer also accepts whole blood, plasma, serum and urine and<br />

has an automatic calibration capability.<br />