05.04.2013 Views

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

MIL-STD 1691 Master Equipment List - The Whole Building Design ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

JSN S0930<br />

Name Washer/Disinfector,STM,1DO,RCSD1WLL,23X19X24 Cham<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army A | Contractor Furnished, Contractor Installed<br />

Navy C | Government Furnished, Government Installed<br />

USAF C | Government Furnished, Government Installed<br />

VA C | Government Furnished, Government Installed<br />

IHS C | Government Furnished, Government Installed<br />

Utilities<br />

NSN 6530<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing F | Cold, hot and treated water and drain<br />

Volts1 208<br />

Utility2: Electrical G | 208/220 volt, 3 phase<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas A | Steam<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc E | Vent to atmosphere<br />

Phase1 3<br />

Phase2 0<br />

Amps1 26<br />

Amps2 0<br />

Weight (lbs) 600 Weight (kgs) 273<br />

Height (in) 76 Height (cen) 193<br />

Width (in) 32 Width (cen) 81<br />

Depth (in) 28 Depth (cen) 71<br />

Volts2 0<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 3500<br />

Watts1 9367<br />

Watts2 0<br />

Description<br />

Unit of Issue EA<br />

Group Sterilizers and Associated <strong>Equipment</strong><br />

A recessed mounted (through one wall), single door (bottom hinged), multi-level, mechanical, washer/disinfector, minimum chamber size<br />

23X19X24, utilizing facility supplied steam. <strong>The</strong> unit is controlled by a microcomputer that monitors and controls all unit operations and<br />

functions and provides both audible and visual indications of deviations. A printer-recorder documents and records each cycle<br />

performance. Processing cycles are pre-programmed with pre-wash, wash/cleaning, rinse, thermal rinse/disinfection, instrument<br />

lubrication and drying. For use in health care facilities and laboratories where reusable items (instruments, utensils, anesthesia/respiratory<br />

goods, and glassware) are handled for decontamination purposes. This unit is not a substitute for terminal sterilization. Unit available<br />

with various electrical configurations, options, and accessories. Price based upon 208V, 3-phase, with various options.<br />

Thursday, April 21, 2011 <strong>MIL</strong>-<strong>STD</strong> <strong>1691</strong> | PPMD<br />

Page 1235 of 1462

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!