JSN L2270<br />

Name Hood, Fume, Radioisotope<br />

Logistical Category (LOGCAT) Fund Source Weights and Measures (Typical Only)<br />

Army A | Contractor Furnished, Contractor Installed<br />

Navy A | Contractor Furnished, Contractor Installed<br />

USAF A | Contractor Furnished, Contractor Installed<br />

VA A | Contractor Furnished, Contractor Installed<br />

IHS A | Contractor Furnished, Contractor Installed<br />

Utilities<br />

NSN 6640<br />

NOTE: Values shown are based on standard typicals used for programming purposes. <strong>The</strong> contractor remains responsible to<br />

develop and coordinate the design / construction in accordance with actual equipment selection.<br />

Utility1: Plumbing B | Cold water and drain<br />

Volts1 120<br />

Utility2: Electrical D | 120 and 208/220 volt<br />

Utility3: Med Gas . |<br />

Utility4: Misc Gas . |<br />

Utility5: Non-Med Gas . |<br />

Utility6: Misc E | Vent to atmosphere<br />

Phase1 1<br />

Phase2 1<br />

Amps1 20<br />

Amps2 10<br />

Weight (lbs) 910 Weight (kgs) 414<br />

Height (in) 96 Height (cen) 244<br />

Width (in) 48 Width (cen) 122<br />

Depth (in) 34 Depth (cen) 86<br />

Volts2 208<br />

Hertz 60<br />

Hertz Dep -1<br />

Hertz Switch 0<br />

BTU / Hour 0<br />

Watts1 2400<br />

Watts2 2080<br />

Description<br />

Unit of Issue EA<br />

Group Laboratory <strong>Equipment</strong><br />

Radioisotope fumehood. <strong>The</strong> unit requires a charcoal trap filter, lead-lined or HEPA filter as required by codes and use criteria in the<br />

exhaust stack, not part of this hood. Other characteristics are that it be strong enough to support lead shielding bricks, have at least two<br />

duplex receptacles and a cup sink. <strong>The</strong> minimum average face velocity shall be 100 fpm and the static pressure loss through the hood<br />

shall not exceed 0.2 inches of water at 100 fpm face velocity (700 CFM).<br />