ESIB<br />

Screen<br />

Marker/Deltamarker<br />

This label displays the position of the last selected marker or<br />

deltamarker in the x and y-directions and the marker/deltamarker index.<br />

As additional information, 2 fields in square brackets are provided which<br />

contain the curve which the marker is assigned to and the active<br />

measuring function of the marker indicated. The measuring function of<br />

the markers in the second field is indicated by the following<br />

abbreviations:<br />

CNT frequency counter active<br />

TRK signal track active<br />

NOI noise measurement active<br />

The marker text has the same color as the trace which the marker is<br />

assigned to. If, for example, the active marker is assigned to trace 1,<br />

and trace 1 is yellow, the marker text will be yellow, too.<br />

Hardware settings<br />

<strong>EMI</strong> Receiver mode<br />

RBW<br />

Det<br />

Meas T<br />

Trd<br />

Unit<br />

Indication of the set resolutíon bandwidth<br />

Indication of the set detectors<br />

Indication of the set measurement time<br />

Indication of the active transducer factors or sets<br />

Indication of unit of measurement results<br />

Analyzer mode<br />

RBW<br />

VBW<br />

SWT<br />

RF Att<br />

Mixer<br />

Unit<br />

Display of the set resolution bandwidth.<br />

Display of the set video bandwidth.<br />

Display of the set sweep time.<br />

Display of the set RF attenuation.<br />

Display of the user set level at the input mixer (only when level differs<br />

from standard settings) ( = level of the reference level REF LVL).<br />

Display of the level unit of the measuring results and the associated<br />

setting and test parameters in full length. This label is particularly<br />

important for selection of units with more than 4 digits, since they can<br />

only be indicated as dB* in the other labels (except for marker).<br />

Tracking generator mode<br />

TG Lvl<br />

Only available with one of the options FSE-B10/11<br />

Indication of the set output level of the tracking generator<br />

Vector Analyzer mode<br />

CF<br />

SR<br />

Demod<br />

Standard<br />

Only available with option FSE-B7<br />

Display of the set center frequency<br />

Display of the symbol rate<br />

Display of the switched on demodulator.<br />

Display of the set standard (e.g. GSM)<br />

Grid<br />

X-axis: frequency or time,<br />

Y-axis: level<br />

1088.7531.12 3.5 E-13