Sweep Control - Analyzer<br />

ESIB<br />

GATE<br />

DELAY<br />

The GATE DELAY softkey activates the entry window for setting the delay<br />

time between the gate signal and the continuation of the sweep.<br />

This may be usefull in allowing for a delay between the gate signal and the<br />

stabilization of an RF carrier.<br />

Gate delay values can be selected between 1 µs and 100 s. The resolution is<br />

dependent upon the absolute value of the delay time:<br />

Gate delay Resolution<br />

0 - 500 µs 1 µs<br />

0.5 - 5 ms 5 µs<br />

5 - 50 ms 50 µs<br />

50 - 500 ms 500 µs<br />

0.5 - 5 s 5 ms<br />

5 - 50 s 50 ms<br />

50 - 100 s 500 ms<br />

In the time domain, a time line is displayed separated from the trigger point by<br />

the gate delay time. This simplifies the adjustment of the necessary delay<br />

time.<br />

The values for GATE DELAY and GATE LENGTH are displayed by two time<br />

lines. The duration of the active sweep for span > 0 (continuation of sweep:<br />

GATE DELAY line, stopping of sweep: GATE LENGTH line) is described by<br />

these two lines. A change in parameters causes a shift of the corresponding<br />

line position. After switching to span > 0, the selected times for the gated<br />

sweep are effective.<br />

IEC-bus command<br />

:[SENSe:]SWEep:EGATe:HOLDoff 100us<br />

GATE<br />

LENGTH<br />

The GATE LENGTH softkey activates the entry window for defining the ESIB<br />

sweep duration.<br />

GATE LENGTH can be set to values between 1µs and 100 s. The resolution<br />

is dependent upon the absolute value of the gate length:<br />

Gate length Resolution<br />

0 - 500 µs 1 µs<br />

0.5 - 5 ms 5 µs<br />

5 - 50 ms 50 µs<br />

50 - 500 ms 500 µs<br />

0.5 - 5 s 5 ms<br />

5 - 50 s 50 ms<br />

50 - 100 s 500 ms<br />

In the time domain (ZERO SPAN), a time line is displayed separated from the<br />

GATE DELAY time by the GATE LENGTH time.<br />

The softkey is only available for the GATE MODE EDGE setting (edge<br />

triggering). It is disabled for the GATE MODE LEVEL setting (level triggering).<br />

IEC/IEEE-bus command :[SENSe:]SWEep:EGATe:LENGth 10ms<br />

GATE<br />

EXTERN<br />

The GATE EXTERN softkey selects a signal applied to the EXT<br />

TRIGGER/GATE connector on the rear panel of the instrument as a gate<br />

source.<br />

IEC/IEEE-bus command<br />

:[SENSe:]SWEep:EGATe:SOURce EXT<br />

GATE<br />

RF POWER<br />

The GATE RF POWER softkey selects the internal RF power detector as a<br />

gate source.<br />

IEC-bus command<br />

:[SENSe:]SWEep:EGATe:SOURce RFPower<br />

1088.7531.12 4.256 E-15