ESIB<br />

Analyzer - Display and Limit Lines<br />

Name and Compatible - Switching on limit lines<br />

A maximum of 8 limit lines can be switched at any one time. A check mark at<br />

the left edge of a cell indicates that this limit line is switched on. A limit line<br />

can only be switched on when it has a check mark in the Compatible column,<br />

i.e., only when the display mode of the x-axis (time or frequency) and y-axis<br />

units are identical to those in the measurement window.<br />

It should be noted that lines with dB scales are compatible with all dB(..)<br />

scales of the Y-axis.<br />

If the trace assigned to a line is not switched on, the line is displayed in the<br />

window the trace would be displayed in.<br />

Example:<br />

In split screen mode, trace 2 is assigned measuring window B. A line<br />

assigned to trace 2 is always displayed in measurement window B.<br />

If the y-axis unit or the domain (frequency or time) are changed, all noncompatible<br />

limit lines are automatically switched off in order to avoid<br />

misinterpretations. The limit lines must be switched on again when the<br />

original display mode is restored.<br />

IEC/IEEE-bus commands<br />

:CALCulate:LIMit:NAME ;<br />

:CALCulate:LIMit:STATe ON | OFF<br />

Limit Check - Activating the limit check<br />

When LIMIT CHECK ON is activated, a limit check is performed automatically<br />

on the active screen. In the center of the screen, a display window containing<br />

the results of the limit check test appears:<br />

LIMIT CHECK: PASSED No out-of-limits conditions.<br />

LIMIT CHECK: FAILED One or more out-of-limits conditions. The<br />

message contains the names of the limit lines<br />

which were violated or whose margins were not<br />

complied with.<br />

LIMIT CHECK: MARGIN<br />

The margin of at least one active limit line was not<br />

complied with, however, no limit line was violated.<br />

The message contains the names of the limit lines<br />

whose margins were not complied with.<br />

The following example shows three active limit lines:<br />

LIMIT CHECK: FAILED<br />

LINE VHF_MASK: Failed<br />

LINE UHF2MASK: Margin<br />

A limit check is performed only if the trace associated with the limit line is<br />

turned on.<br />

If LIM CHECK is set to OFF for all active limit lines, then the limit line check is<br />

not executed and the display field not activated.<br />

IEC/IEEE-bus command :CALCulate:LIMit:FAIL?<br />