Transmission measurement - Tracking generator<br />

ESIB<br />

If, after source calibration, a 10 dB attenuator is inserted into the signal<br />

path between DUT and analyzer input, the measurement trace will be<br />

moved down by 10 dB. Entering a REF VALUE of –10 dB then moves the<br />

reference line down 10 dB. This means that the measured trace and the<br />

reference trace coincide again.<br />

After the reference line has been moved by entering a REF VALUE of –10<br />

dB, deviations from the nominal power level can then be displayed with a<br />

high resolution (e.g. 1 dB/div). The power is still displayed in absolute<br />

values, which means that in the above example 1 dB below the nominal<br />

power (reference line) is equivalent to an attenuation of 11 dB.<br />

Fig. 4-25<br />

Measurement of a 10dB attenuator pad with 1 dB / Div<br />

IEC/IEEE bus command<br />

:DISPlay[:WINDow]:TRACe:Y[:SCALe]:RVALue<br />

<br />

RECALL<br />

The RECALL softkey restores the instrument setting that were used during<br />

calibration.<br />

This can be useful if the device setting was changed after calibration (eg<br />

center frequency setting, frequency deviation, reference level, etc. ).<br />

The softkey is only offered if:<br />

• the analyzer mode has been selected<br />

• the memory contains a calibration data set.<br />

IEC/IEEE bus command :[SENSe:]CORRection:RECall<br />

1088.7531.12 4.270 E-15