Initial Configuration<br />

ESIB<br />

General Configuration – SYSTEM and CONFIGURATION Key<br />

Groups<br />

ESIB Initial Configuration – PRESET Key<br />

SYSTEM<br />

PRESET CAL<br />

DI AY INFO<br />

Using the PRESET key, the ESIB can be brought to a predefined initial state<br />

equivalent to the state after power on. All previous setting are deleted if they<br />

were not previously saved in memory. Returning to the PRESET state is not<br />

critical with regard to an arbitrary signal which may still be present at the input<br />

as long as the signal is within the specified voltage range.<br />

Note:<br />

The initial state set by the PRESET key can be adapted to the<br />

user requirements using the AUTO RECALL function. When the<br />

PRESET key is pressed, the AUTO RECALL dataset is loaded.<br />

For further information refer to Section "Saving and Recalling<br />

Data Sets".<br />

IEC/IEEE- bus command<br />

:*RST<br />

Pressing the PRESET key, causes the ESIB to enter its initial state according to the following table:<br />

Table 4-1<br />

Parameter<br />

Initial state of the ESIB<br />

Setting<br />

Mode<br />

<strong>EMI</strong> Receiver<br />

Receiver Frequency<br />

100 MHz<br />

RF Attenuation<br />

Auto<br />

Preamplification<br />

off<br />

Input Input 1<br />

Detector<br />

AV<br />

Measurement Time<br />

100 ms<br />

IF Bandwidth (RES BW)<br />

120 kHz<br />

Demodulation<br />

off<br />

Trigger<br />

free run<br />