Traces - Analyzer<br />

ESIB<br />

TRACE MATH<br />

T1-T2+REF<br />

-> T1<br />

T1-T3+REF<br />

-> T1<br />

T1-T4+REF<br />

-> T1<br />

T1-REF<br />

->T1<br />

The T1-T2+REF, T1-T3+REF, T1-T3+REF and T1-REF softkeys subtract the<br />

traces and add the set level to the difference. If the reference line is switched<br />

on (see key D LINES), the level value of the reference line instead of the<br />

reference level is added to the difference. Thus, the differential curve can be<br />

positioned anywhere on the screen by shifting the reference line. The<br />

difference of the two traces with respect to the reference line is displayed.<br />

The T1-REF softkey subtracts the level of the reference line from the trace<br />

level.<br />

To indicate that the trace is a difference trace, a enhancement label is<br />

displayed at the right margin of the measurement value diagram (1-2, 1-3, 1-<br />

4, 1-R). In TRACE 1 main menu, the TRACE MATH softkey is on a coloured<br />

background to show that the function is being used.<br />

Important: For a display with two measurement windows, not all<br />

combinations are allowed if the sweep data for screen A and<br />

screen B are not the same. Only the traces allowed in the screen<br />

can be combined (in screen A, only trace 1 with trace 3; in<br />

screen B, only trace 2 with trace 4).<br />

IEC/IEEE-bus command<br />

:CALCulate:MATH:STATe ON<br />

:CALCulate:MATH[:EXPRession][:DEFine] <br />

TRACE MATH<br />

OFF<br />

The TRACE MATH OFF softkey switches the math function off. The softkey<br />

is only available when a conversion function is enabled.<br />

IEC/IEEE-bus command :CALCulate:MATH:STATe OFF<br />

ADJUST TO<br />

TRACE<br />

The ADJUST TO TRACE softkey restores the original instrument settings<br />

when the corresponding trace is set to VIEW and the current instrument<br />

settings are different than those of the trace data.<br />

If a trace is frozen with VIEW, changes in the instrument settings can be<br />

made without influencing the trace. An asterisk( * ) at the edge of the display<br />

screen indicates that the current instrument settings are different than those<br />

in effect when the trace was recorded. In this case, the ADJUST TO TRACE<br />

softkey is presented with which the original instrument setting can be<br />

restored.<br />

IEC/IEEE-bus command --<br />

1088.7531.12 4.236 E-15