ESIB<br />

Analyzer - Sweep Control<br />

SWEEP SWEEP-GATE SETTINGS sub-menu:<br />

GATE<br />

SETTINGS<br />

GATE<br />

SETTINGS<br />

GATE<br />

LEVEL<br />

GATE MODE<br />

LEVEL EDGE<br />

GATE POL<br />

POS NEG<br />

GATE<br />

DELAY<br />

GATE<br />

LENGTH<br />

In the GATE SETTINGS submenu, all settings are made<br />

which are necessary for ’gated sweep’ operation.<br />

On switching to the time domain, the GATE DELAY and<br />

GATE LENGTH times are represented by horizontal time<br />

lines which allow simple adjustment of the gate time.<br />

The softkeys GATE EXTERN and GATE RF POWER provide<br />

alternative settings, only one of them can be active at one<br />

time.<br />

GATE<br />

EXTERN<br />

GATE<br />

RF POWER<br />

GATE<br />

ADJUST<br />

GATE<br />

LEVEL<br />

The GATE LEVEL softkey activates the entry window for defining the<br />

threshold value of the external gate signal.<br />

The threshold can be set between -5V and +5V.<br />

IEC/IEEE-bus command :[SENSe:]SWEep:EGATe:LEVel 3V<br />

GATE MODE<br />

LEVEL EDGE<br />

The GATE MODE LEVEL/EDGE softkey selects the trigger mode. The sweep<br />

mode GATE can be operated either as level or as edge triggered.<br />

For level triggering, the GATE LENGTH softkey is deactivated and can not be<br />

used.<br />

IEC/IEEE-bus command<br />

:[SENSe:]SWEep:EGATe:TYPE LEVel | EDGE<br />

GATE POL<br />

POS NEG<br />

The GATE POL softkey determines the polarity of the GATE control line.<br />

If level triggering with GATE POL POS is set, the sweep is stopped when the<br />

gate signal is a logic ´ 0 ' (input signal < gate level) at the EXT<br />

TRIGGER/GATE input and then, when the gate signal goes to a logic ' 1 ', the<br />

sweep is continued after a delay time of GATE DELAY.<br />

In the case of edge triggering, the sweep is continued on a ' 0 ' to ' 1'<br />

transition for the duration of GATE LENGTH after a delay (GATE DELAY) has<br />

elapsed.<br />

IEC/IEEE-bus command<br />

:[SENSe:]SWEep:EGATe:POLarity POS|NEG<br />

1088.7531.12 4.255 E-15