ESIB<br />

DISPlay Subsystem<br />

:DISPlay[:WINDow]:TIME ON | OFF<br />

This command switches on or off the screen display of date and time.<br />

Example: ":DISP:TIME ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

The numeric suffix in WINDow is not significant.<br />

:DISPlay[:WINDow]:TRACe:X[:SCALe]:RVALue <br />

This command defines the reference value for the X-axis of the measuring diagram.<br />

Example: ":DISP:TRAC:X:RVAL 20SYM"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

VA-D<br />

:DISPlay[:WINDow]:TRACe:X[:SCALe]:ZOOM ON | OFF<br />

This command displays the zoomed current frequency range in the opposite window of the split<br />

screen representation.<br />

Example: ":DISP:TRAC:X:ZOOM ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

A-F<br />

The numeric suffix in TRACe is not significant.<br />

:DISPlay[:WINDow]:TRACe:X[:SCALe]:ZOOM[:FREQuency]:STARt<br />

<br />

This command defines the start frequency of the zoomed display range. The value must lie between<br />

start and stop frequency of the original range.<br />

Example: ":DISP:TRAC:X:ZOOM:STAR 100MHZ"<br />

Features: *RST value: -- (depending on the current frequency setting)<br />

SCPI: device-specific<br />

Mode:<br />

A-F<br />

The numeric suffix in TRACe is not significant.<br />

:DISPlay[:WINDow]:TRACe:X[:SCALe]:ZOOM[:FREQuency]:STOP <br />

This command defines the stop frequency of the zoomed display range. The value must lie between<br />

start and stop frequency of the original range.<br />

Example: ":DISP:TRAC:X:ZOOM:STOP 200MHZ"<br />

Features: *RST value: -- (depending on the current frequency setting)<br />

SCPI: device-specific<br />

Mode:<br />

A-F<br />

The numeric suffix in TRACe is not significant.<br />