ESIB<br />

DISPlay Subsystem<br />

:DISPlay[:WINDow]:TRACe:MODE WRITe | VIEW | AVERage | MAXHold | MINHold |<br />

FRESult<br />

This command defines the type of display and the evaluation of the traces.<br />

For selection FRESult (display of final measurement) only suffix 3 or 4 is valid for TRACe.<br />

Example: ":DISP:TRAC3:MODE MAXH"<br />

Features: *RST value: WRITe for TRACe1, STATe OFF for TRACe2 to 4<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

:DISPlay[:WINDow]:TRACe:MODE:CWRite ON | OFF<br />

This command selects continuous display of the measured values (continuous write).<br />

Example: ":DISP:TRAC3:MODE:CWR ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

VA<br />

:DISPlay[:WINDow]:TRACe:MODE:ANALog ON | OFF<br />

This command selects continuous display of the measured values in the analyzer mode (analog<br />

trace).<br />

Example: ":DISP:TRAC3:MODE:ANAL ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

:DISPlay[:WINDow]:TRACe:MODE:HCONtinuous ON | OFF<br />

This command defines whether the traces in the min hold/max hold mode are reset after some<br />

definite parameter changes.<br />

Example: ":DISP:TRAC3:MODE:HCON ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

OFF<br />

The traces are reset after some definite parameter changes.<br />

ON<br />

This mechanism is switched off.<br />

In general, parameter changes require a restart of the measurement before results are evaluated<br />

(e.g. with markers). For those changes that are known to require a new measurement (e.g.<br />

modification of the span), the trace is automatically reset so that erroneous evaluations of previous<br />

results are avoided. This mechanism can be switched off for those exceptional cases where the<br />

described behavior is unwelcome.<br />