14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ESIB<br />

SENSe Subsystem<br />

SENSe:SCAN - Subsystem<br />

This subsystem controls the parameters for the receiver scan data. The numeric suffix in<br />

[SENSe]SCAN is not significant.<br />

COMMAND PARAMETER UNIT COMMENT<br />

[SENSe]<br />

:SCAN<br />

:STARt<br />

:STOP<br />

:STEP<br />

:BANDwidth<br />

:RESolution<br />

:TIME<br />

:INPut<br />

:TYPe<br />

:ATTenuation<br />

:AUTO<br />

:GAIN<br />

:STATe<br />

:AUTO<br />

:RANGes<br />

[:COUNt]<br />

<br />

<br />

<br />

<br />

<br />

INPUT1 | INPUT2<br />

<br />

<br />

<br />

<br />

<br />

HZ<br />

HZ<br />

HZ<br />

HZ<br />

s<br />

dB<br />

--<br />

--<br />

--<br />

--<br />

:[SENSe:]SCAN:STARt f min to f max<br />

This command defines the start frequency of a receiver subscan.<br />

Example: ":SCAN1:STAR 50kHz"<br />

Features: *RST value: 150 kHz (range1)<br />

30 MHz (range2)<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:STOP f min to f max<br />

This command defines the stop frequency of a receiver subscan.<br />

Example: ":SCAN1:STOP 50kHz"<br />

Features: *RST value: 30 MHz (range1)<br />

1 GHz (range2)<br />

SCPI: device-specific<br />

Mode:<br />

R<br />

:[SENSe:]SCAN:STEP f min to f max<br />

This command defines the step size for the frequency of a receiver subscan.<br />

Example: ":SCAN1:STEP 100Hz"<br />

Features: *RST value: 4 kHz (range 1)<br />

40 kHz (range 2)<br />

SCPI: device-specific<br />

Mode:<br />

R<br />