Level Display/RF Input - Analyzer<br />

ESIB<br />

ATTEN AUTO<br />

NORMAL<br />

The ATTEN AUTO NORMAL softkey sets the RF attenuation automatically as<br />

a function of the selected reference level.<br />

IEC/IEEE-bus commands<br />

:INPut:ATTenuation:AUTO:MODE NORMal<br />

:INPut:ATTenuation:AUTO ON<br />

ATTEN AUTO<br />

LOW NOISE<br />

The ATTEN AUTO LOW NOISE softkey sets the RF attenuation always 10dB<br />

lower than in the RF ATTEN AUTO mode. Therefore, for 10 dB RF<br />

attenuation, the maximum reference level is -10 dBm. For reference levels<br />

which are lower, at least 10 dB is always set (see above).<br />

Setting Low Noise means that the indicated inherent noise level is low. This<br />

setting is recommended if signals with a low level have to be measured as<br />

the highest S/N ratio is then obtained.<br />

IEC/IEEE-bus commands<br />

:INPut:ATTenuation:AUTO:MODE LNOise<br />

:INPut:ATTenuation:AUTO ON<br />

ATTEN AUTO<br />

LOW DIST<br />

The ATTEN AUTO LOW DIST softkey sets the RF attenuation 10 dB higher<br />

than in RF ATT AUTO mode. This means that for 10 dB RF attenuation, the<br />

maximum reference level is -30 dBm (-40 dBm at the mixer).<br />

This setting is recommended whenever a small signal is to be measured in<br />

the presence of larger signals. Here, the intermodulation-free range of the<br />

ESIB is large and the internal distortion products are minimised.<br />

IEC/IEEE-bus commands<br />

:INPut:ATTenuation:AUTO:MODE LDIStortion<br />

:INPut:ATTenuation:AUTO ON<br />

MIXER<br />

LEVEL<br />

The MIXER LEVEL softkey activates the entry of the maximum mixer level<br />

attainable for a given reference level. At the same time it enables softkeys<br />

ATTEN AUTO LOW NOISE and ATTEN AUTO LOW DIST .<br />

Range of input values is -10 to -100 dBm<br />

IEC/IEEE-bus command :INPut:MIXer -30DBM<br />

1088.7531.12 4.172 E-15