CALCulate Subsystem<br />

:CALCulate:LIMit:ACPower[:STATe] ON | OFF<br />

This command switches on and off the limit check for adjacent channel power measurements. The<br />

commands CALC:LIM:ACP:ACH:STAT or CALC:LIM:ACP:ALT:STAT must be used in addition to<br />

specify whether the limit check is to be performed for the upper/lower adjacent channel or for the<br />

alternate adjacent channels.<br />

Examples:<br />

":CALC:LIM:ACP ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

The numeric suffixes or are not significant for this command.<br />

:CALCulate:LIMit:ACPower:ACHannel 0 to 100 dB, 0 to 100 dB<br />

This command defines the limit for the upper/lower adjacent channel for adjacent channel power<br />

measurements.<br />

Parameter:<br />

Examples:<br />

The first (second) numeric value is the limit for the upper (lower) adjacent<br />

channel.<br />

":CALC:LIM:ACP:ACH 30DB, 30DB"<br />

Features: *RST value: 0 dB<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

The numeric suffixes or are not significant for this command.<br />

:CALCulate:LIMit:ACPower:ACHannel:STATe ON | OFF<br />

This command activates the limit check for the adjacent channel when adjacent channel power<br />

measurement is performed. Before, the limit check must be activated using CALC:LIM:ACP ON.<br />

Examples:<br />

":CALC:LIM:ACP:ACH:STAT ON"<br />

Features: *RST value: OFF<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

The numeric suffixes or are not significant for this command.<br />

:CALCulate:LIMit:ACPower:ACHannel:RESult?<br />

This command queries the result of the limit check for the upper/lower adjacent channel when<br />

adjacent channel power measurement is performed.<br />

Parameter: The result is returned in the form , where<br />

= PASSED | FAILED, and where the first returned value denotes the<br />

lower, the second denotes the upper adjacent channel.<br />

Examples: ":CALC:LIM:ACP:ACH:RES?"<br />

Features: *RST value: --<br />

SCPI: device-specific<br />

Modes:<br />

A, VA<br />

This command is a query and therefore not assigned a *RST value. If the power measurement of the<br />

adjacent channel is switched off, the command triggers a query error.<br />