SENSe Subsystem<br />

ESIB<br />

:[SENSe:]CORRection:CVL:MIXer <br />

This command defines the type designation of the mixer in the Conversion Loss Table.<br />

Parameter: ::= Type designation of mixer with a maximum of 16 characters<br />

Example: ":CORR:CVL:MIX ’FSE_Z60’"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid<br />

in conjunction with option External Mixer Output, FSE-B21.<br />

:[SENSe:]CORRection:CVL:SNUMber <br />

This command defines the serial number of the mixer in the Conversion Loss Table.<br />

Parameter: ::= Serial number of mixer with a maximum of 16 characters<br />

Example: ":CORR:CVL:SNUM ’123.4567’"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid<br />

in conjunction with option External Mixer Output, FSE-B21.<br />

:[SENSe:]CORRection:CVL:BAND A|Q|U|V|E|W|F|D|G|Y|J<br />

This command defines the waveguide band in the Conversion Loss Table.<br />

Example: ":CORR:CVL:BAND E"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid<br />

in conjunction with option External Mixer Output, FSE-B21.<br />

:[SENSe:]CORRection:CVL:TYPE ODD | EVEN | EODD<br />

This command defines the type of harmonic in the Conversion Loss Table.<br />

Example: ":CORR:CVL:TYPE EODD"<br />

Features: *RST value: -<br />

SCPI: device-specific<br />

Mode:<br />

A<br />

Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid<br />

in conjunction with option External Mixer Output, FSE-B21.<br />