SENSe Subsystem<br />

ESIB<br />

SENSe:ROSCillator Subsystem<br />

This subsystem controls the reference oscillator. The suffix in SENSe is not significant in this<br />

subsystem.<br />

COMMAND PARAMETERS UNIT COMMENT<br />

[SENSe]<br />

:ROSCillator<br />

:SOURce<br />

:EXTernal<br />

:FREQuency<br />

INTernal | EXTernal<br />

<br />

--<br />

HZ<br />

[:INTernal]<br />

:TUNe<br />

<br />

:SAVe<br />

no query<br />

:[SENSe:]ROSCillator:SOURce<br />

INTernal | EXTernal<br />

This command controls selection of the reference oscillator.<br />

Example: ":ROSC:SOUR EXT"<br />

Features: *RST value: –<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

If the external reference oscillator is selected, the reference signal must be connected to the rear<br />

panel of the instrument. *RST has no influence on this setting.<br />

:[SENSe:]ROSCillator:EXTernal:FREQuency 1MHz to 16MHz<br />

This command defines the frequency of the external reference oscillator.<br />

Example: ":ROSC:EXT:FREQ 5MHz"<br />

Features: *RST value: 10MHz<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

The value of the external reference frequency (1MHz to 16MHz) is rounded in steps of 1MHz.<br />

:[SENSe:]ROSCillator[:INTernal]:TUNe 0 to 4095<br />

This command defines the value for the tuning of the internal reference oscillator.<br />

Example: ":ROSC:TUN 128"<br />

Features: *RST value: –<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

The reference oscillator should be tuned only if an error has been detected in the frequency accuracy<br />

check. After PRESET or switching on the instrument, the saved value of the reference frequency is<br />

restored.<br />

:[SENSe:]ROSCillator[:INTernal]:TUNe:SAVe<br />

This command saves the new value for the tuning of the internal reference oscillator. The factory-set<br />

value in the EEPROM is overwritten.<br />

Example: ":ROSC:TUN:SAV"<br />

Features: *RST value: –<br />

SCPI: device-specific<br />

Modes:<br />

R, A, VA<br />

1088.7531.12 6.146 E-15