ESIB<br />

Receiver - Traces<br />

AVERAGE<br />

The AVERAGE softkey activates the average detector.<br />

The average detector supplies the straight average of the<br />

signal during the measurement time.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector[:FUNCtion] AVERage<br />

RMS<br />

The RMS softkey activates the rms detector.<br />

The rms detector supplies the rms value of the signal. The<br />

mean of squares of all sampled values is formed during the<br />

measurement time.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector[:FUNCtion] RMS<br />

MIN PEAK<br />

The MIN PEAK softkey activates the min peak detector.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector[:FUNCtion] NEGative<br />

AC VIDEO<br />

The DETECTOR AC VIDEO softkey activates the AC VIDEO<br />

detector. The softkey is available only if the instrument is<br />

equipped with the linear video output option (option ESIB-B1).<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector[:FUNCtion] ACVideo<br />

FINAL<br />

MAX PEAK<br />

The FINAL MAX PEAK selects the max peak detector for the<br />

final measurement.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector:FMEasurement POSitive<br />

FINAL<br />

QUASIPEAK<br />

The FINAL QUASIPEAK selects the quasi peak detector for<br />

the final measurement.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector:FMEasurement QPEak<br />

FINAL<br />

AVERAGE<br />

The FINAL AVERAGE selects the average detector for the<br />

final measurement.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector:FMEasurement AVERage<br />

FINAL<br />

RMS<br />

The FINAL RMS selects the rms detector for the final<br />

measurement.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector:FMEasurement RMS<br />