Traces - Receiver<br />

ESIB<br />

In SINGLE SCAN mode, RUN SCAN initiates n single scans. The scans are stopped as soon as the<br />

selected number of scans is reached.<br />

SCAN<br />

COUNT<br />

The SCAN COUNT softkey activates the entry of the number of scans for<br />

SINGLE SCAN.<br />

The permitted range for SCAN COUNT is 1 through 32767. The default setting is<br />

1 scan.<br />

IEC\IEEE bus command :[SENSe:]SWEep:COUNt 10<br />

DETECTOR<br />

DETECTOR<br />

MAX PEAK<br />

QUASIPEAK<br />

AVERAGE<br />

RMS<br />

MIN PEAK<br />

FINAL<br />

MAX PEAK<br />

FINAL<br />

QUASIPEAK<br />

FINAL<br />

AVERAGE<br />

FINAL<br />

RMS<br />

FINAL<br />

MIN PEAK<br />

The DETECTOR softkey opens a submenu to select the<br />

detector.<br />

The detector type may be indepently selected for each trace.<br />

Multiple detection is activated by switching on up to four single<br />

detectors.<br />

For description of detector types see Section "Detector<br />

Selection" in Analyzer Mode.<br />

The detectors for the final measurement are marked by<br />

"FINAL".<br />

Softkey AC VIDEO is available only if the instrument is<br />

equipped with the linear video output option (option ESIB-B1).<br />

It replaces softkey MIN PEAK.<br />

MAX PEAK<br />

The MAX PEAK softkey activates the max peak detector.<br />

IEC\IEEE bus command<br />

:[SENSe:]DETector[:FUNCtion] POSitive<br />

QUASIPEAK<br />

The QUASIPEAK softkey selects the quasi-peak detector.<br />

The quasi-peak detector yields the maximum detected value<br />

weighted to CISPR 16 during the set measurement time. The<br />

IF bandwidth is adapted as a function of the frequency range.<br />

The coupling of the IF bandwidth to the frequency range can<br />

be cancelled using softkey QP RBW UNCOUPLED in the <strong>EMI</strong><br />

<strong>RECEIVER</strong> DETECTOR menu.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector[:FUNCtion] QPEak<br />

1088.7531.12 4.146 E-15