ESIB<br />

ABORt / CALCulate Subsystem<br />

ABORt Subsystem<br />

The ABORt subsystem contains the commands for aborting triggered actions. An action can be<br />

triggered again immediately after being aborted. All commands trigger events which is why they are not<br />

assigned any *RST value.<br />

ABORt<br />

HOLD<br />

COMMAND PARAMETERS UNIT COMMENT<br />

--<br />

--<br />

--<br />

--<br />

no query<br />

no query<br />

:ABORt<br />

This command aborts a current measurement and resets the trigger system.<br />

Example: "ABOR;INIT:IMM"<br />

Features: *RST value: 0<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

:HOLD<br />

This command interrupts a current scan measurement.<br />

Example: "HOLD"<br />

Features: *RST value: -<br />

SCPI: conforming<br />

Modes:<br />

R<br />

CALCulate Subsystem<br />

The CALCulate subsystem contains commands for converting instrument data, transforming and<br />

carrying out corrections. These functions are carried out subsequent to data acquistion, i.e., following<br />

the SENSe subsystem.<br />

In the split-screen representation, a distinction is made between CALCulate1 and CALCulate2:<br />

CALCulate1 =;^ screen A;<br />

CALCulate2 =;^ screen B<br />

For setting REAL/IMAG PART in Vector Analyzer mode a distinction is also made between CALCulate3<br />

and CALCulate4 in the split-screen representation:<br />

CALCulate3 =;^ screen C;<br />

CALCulate4 =;^ screen D<br />

1088.7531.12 6.7 E-15