Display Configuration<br />

ESIB<br />

Measurement Window Coupling<br />

For the most part, the settings for the two windows can be either independently performed or coupled to<br />

one another. In many cases, when changes are made in one window (e.g., the reference level) it is<br />

desirable to also make the corresponding changes in the second window. The measurement window<br />

parameters which are to be coupled may be selected in the SCREEN COUPLING menu.<br />

SYSTEM DISPLAY-SCREEN COUPLING submenu:<br />

SCREEN<br />

COUPLING<br />

SCREEN<br />

COUPLING<br />

MODE<br />

COUPLED<br />

The SCREEN COUPLING softkey calls a sub-menu in<br />

which the coupling between the two measurement<br />

windows SCREEN A and SCREEN B can be defined. This<br />

coupling is only effective when both windows are<br />

displayed (SPLIT SCREEN).<br />

HORIZONTAL<br />

SCALING<br />

In the default setting, all of the items which are selectable<br />

are also coupled.<br />

VERTICAL<br />

SCALING<br />

COUPLING<br />

CONTROL<br />

...<br />

SCREENS<br />

UNCOUPLED<br />

MODE<br />

COUPLED<br />

The MODE COUPLED softkey switches the coupling of the operating mode<br />

(Analyzer, Vector Analyzer) on/off.<br />

IEC/IEEE- bus command :INSTrument:COUPle MODE<br />

HORIZONTAL<br />

SCALING<br />

The HORIZONTAL SCALING softkey switches the coupled scaling of the<br />

horizontal axes on and off. In the frequency domain, the center frequency and<br />

the frequency span are identical. In the time domain, the sweep time for both<br />

windows is the same.<br />

IEC/IEEE- bus command :INSTrument:COUPle X<br />

VERTICAL<br />

SCALING<br />

COUPLING<br />

CONTROL<br />

The VERTICAL SCALING softkey switches the coupled scaling of the vertical<br />

axes on and off. For a level measurement, this means that the reference level<br />

and the vertical resolution (LEVEL RANGE) are set to the same values in<br />

both windows.<br />

IEC/IEEE- bus command : INSTrument:COUPle Y<br />

The COUPLING CONTROL softkey switches the coupling of the trigger and<br />

gate parameters and of SWEEP COUNT and SWEEP SINGLE/<br />

CONTINOUS on/off.<br />

IEC/IEEE- bus command :INSTrument:COUPle CONTrol<br />

SCREENS<br />

UNCOUPLED<br />

The SCREENS UNCOUPLED softkey switches all of the possible couplings<br />

between the measurement windows to off.<br />

IEC/IEEE- bus command : INSTrument:COUPle NONE | ALL<br />