14.06.2014 Views

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

Operating Manual EMI TEST RECEIVER ESIB7 - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ESIB- Receiver<br />

ESIB<br />

CONFIGURATION MODE - <strong>EMI</strong> <strong>RECEIVER</strong> menu:<br />

DETECTOR<br />

DETECTOR<br />

MAX PEAK<br />

QUASIPEAK<br />

AVERAGE<br />

RMS<br />

The DETECTOR softkey opens a submenu to select the<br />

detector.<br />

Multiple detection is activated by switching on up to four single<br />

detectors. The MIN PEAK, RMS and AC VIDEO detector<br />

cannot be switched on simultaneously.<br />

Softkey AC VIDEO is available only if the instrument is<br />

equipped with the linear video output (option ESIB-B1).<br />

MIN PEAK<br />

AC VIDEO<br />

QP RBW<br />

COUPLED<br />

MAX PEAK<br />

The MAX PEAK softkey activates the max peak detector.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector:RECeiver[:FUNCtion] POSitive<br />

QUASIPEAK<br />

The QUASIPEAK softkey selects the quasi-peak detector.<br />

The quasi-peak detector yields the maximum detected value<br />

weighted to CISPR 16 during the set measurement time. The IF<br />

bandwidth is adapted as a function of the frequency range. The<br />

coupling of the IF bandwidth to the frequency range can be<br />

cancelled using softkey QP RBW UNCOUPLED.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector:RECeiver[:FUNCtion] QPEak<br />

AVERAGE<br />

The AVERAGE softkey activates the average detector.<br />

The average detector supplies the straight average of the signal<br />

during the measurement time.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector:RECeiver[:FUNCtion] AVERage<br />