Traces - Analyzer<br />

ESIB<br />

DETECTOR<br />

SAMPLE<br />

The DETECTOR SAMPLE softkey selects the sample detector.<br />

It is used for measuring uncorrelated signals such as noise. The power can<br />

be determined by means of fixed correction factors for evaluation and the log<br />

amplifier.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector SAMPle<br />

DETECTOR<br />

RMS<br />

The DETECTOR RMS softkey activates the rms detector.<br />

The rms detector gives the true power of the signal irrespective of the<br />

waveform. The mean of the squares of all level samples is formed as a pixel<br />

is swept. The sweep time, therefore, determines the number of values<br />

averaged and, with increasing sweep time, better averaging is obtained. The<br />

rms detector is therefore an alternative to averaging over several sweeps<br />

(see TRACE AVERAGE).<br />

In the time domain (SPAN = 0), the rms detector is only available for sweep<br />

times ≥ 5 ms. Moreover, the combination of the rms detector with the<br />

pretrigger function and the gaped sweep function is not permissible.<br />

The video bandwidth has to be set to at least 10 times the resolution<br />

bandwidth (RBW) so that the rms value of the signal is not invalidated by<br />

video filtering.<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector RMS<br />

DETECTOR<br />

AVERAGE<br />

The DETECTOR AVERAGE softkey activates the average detector.<br />

In contrast to the rms detector, the average detector supplies the straight<br />

average of all the level sampleds as a pixel is swept over.<br />

The same restrictions as those of the rms detector apply (see above).<br />

IEC/IEEE-bus command<br />

:[SENSe:]DETector AVERage<br />

DETECTOR<br />

AC VIDEO<br />

The DETECTOR AC VIDEO softkey activates the AC VIDEO detector.<br />

Irrespective of the signal form, the AC VIDEO detector always supplies the<br />

AC component of the signal. For this purpose the difference is formed of all<br />

maximum and minimum level values collected during the runtime of the pixel<br />

/ the set measurement time. The sweep / measurement time therefore<br />

determines the number of values to be considered, so the AC component is<br />

determined more accurately with increasing sweep / measurement time. The<br />

AC VIDEO detector can thus be used as an alternative for the detection of<br />

modulated signals.<br />

The softkey is available only if the instrument is equipped with the linear video<br />

output (option ESIB-B1).<br />

IEC/IEEE-bus command :[SENSe:]DETector ACVideo<br />