SENSe Subsystem<br />

ESIB<br />

SENSe:AVERage Subsystem<br />

The SENSe:AVERage subsystem calculates the average of the data acquired. A new test result is<br />

obtained from various successive measurements. The amount of test points and the axis reference of<br />

the new result correspond to those of the original measurements.<br />

[SENSe]<br />

:AVERage<br />

:COUNt<br />

:AUTO<br />

[:STATe]<br />

:TYPE<br />

COMMAND PARAMETERS UNIT COMMENT<br />

<br />

<br />

<br />

MAXimum | SCALar<br />

--<br />

--<br />

--<br />

--<br />

:[SENSe:]AVERage:COUNt 0 to 32767<br />

The command specifies the number of measurements which are combined.<br />

Example: ":AVER:COUN 16"<br />

Features: *RST value: 0<br />

SCPI: conforming<br />

Modes:<br />

R, A. VA-D<br />

:[SENSe:]AVERage:COUNt:AUTO ON | OFF<br />

AUTO ON selects a suitable number of :COUNt for the respective type of measurement.<br />

Example: ":AVER:COUN:AUTO ON"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Modes:<br />

R, A. VA-D<br />

:[SENSe:]AVERage[:STATe] ON | OFF<br />

The command switches on or off the average function.<br />

Example: "AVER OFF"<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Modes:<br />

R, A, VA<br />

1088.7531.12 6.104 E-15