SENSe Subsystem<br />

ESIB<br />

:[SENSe:]CORRection[:STATe] ON | OFF<br />

This command activates/deactivates normalization of the measurement results.<br />

Example: ":CORR ON "<br />

Features: *RST value: OFF<br />

SCPI: conforming<br />

Mode:<br />

A<br />

This command is only valid in conjunction with option Tracking Generator.<br />

:[SENSe:]CORRection:METHod TRANsmission | REFLexion<br />

This command selects the kind of measurement with active tracking generator<br />

(transmission/reflexion).<br />

Example: ":CORR:METH TRAN "<br />

Features: *RST value: TRANsmission<br />

SCPI: device specific<br />

Mode:<br />

A<br />

This command is only valid in conjunction with option Tracking Generator.<br />

:[SENSe:]CORRection:COLLect[:ACQuire] THRough | OPEN<br />

This command selects the kind of measurement for the reference values of the normalization<br />

(response calibration).<br />

Example: ":CORR:COLL THR"<br />

Features:<br />

Mode:<br />

THRough<br />

OPEN<br />

*RST value:<br />

SCPI:<br />

conforming<br />

A<br />

"TRANsmission" mode: calibration with direct connection between tracking<br />

generator and device input.<br />

"REFLexion" mode: calibration with short circuit at the input<br />

only valid in "REFLexion" mode:calibration with open input<br />

This command is an event which is why it is not assigned an *RST value an a query. It is only valid in<br />

conjunction with option Tracking Generator.<br />

:[SENSe:]CORRection:RECall<br />

This command restores the instrument setting that was valid for the measurement of the reference<br />

data.<br />

Example: ":CORR:REC"<br />

Features: *RST value: -<br />

SCPI: conforming<br />

Mode:<br />

A<br />

This command is an event which is why it is not assigned an *RST value and a query. It is only valid<br />

in conjunction with option Tracking Generator.<br />