ESIB- Receiver<br />

ESIB<br />

HOLD<br />

FINAL MEAS<br />

CONTINUE<br />

FINAL MEAS<br />

AUTOMATIC<br />

FINAL<br />

INTER<br />

ACTIVE<br />

SKIP<br />

FREQUENCY<br />

MEASURE<br />

STOP<br />

FINAL MEAS<br />

The HOLD FINAL MEAS softkey interrupts the<br />

automatic run of the final measurement.<br />

The CONTINUE FINAL MEAS submenu<br />

appears.<br />

With the final measurement halted all receiver<br />

settings can be modified for example for<br />

examining the signal in detail.<br />

The mode of the measurement (automatic or<br />

interactive) can again be selected.<br />

The final measurement can be started with the<br />

MEASURE softkey.<br />

IEC/IEEE-bus command --<br />

AUTOMATIC<br />

FINAL<br />

INTER<br />

ACTIVE<br />

SKIP<br />

FREQUENCY<br />

The AUTOMATIC FINAL softkey selects the<br />

automatic mode for the final measurement<br />

(see above).<br />

IEC/IEEE-bus command --<br />

The INTERACTIVE softkey selects the<br />

interactive mode for the final measurement as<br />

described above.<br />

IEC/IEEE-bus command --<br />

The SKIP FREQUENCY softkey skips the next<br />

entry in the FREQUENCY column.<br />

IEC/IEEE-bus command --<br />

MEASURE<br />

The MEASURE softkey continues with the final<br />

measurement. The final measurement starts at<br />

the next frequency entry on the peak list or at a<br />

marked frequency if one or several lines have<br />

been skipped with the SKIP FREQUENCY<br />

softkey.<br />

IEC/IEEE-bus command --<br />

STOP<br />

FINAL MEAS<br />

The STOP FINAL MEAS softkey halts the final measurement.<br />

The final measurement starts from the beginning on a new start.<br />

The previously collected data are lost.<br />

IEC/IEEE-bus command --<br />