ESIB<br />

Configuration/Setup<br />

Preliminary Setup and Interface Configuration – SETUP Key<br />

CONFIGURATION SETUP menu:<br />

CONFIGURATION<br />

MODE<br />

SETUP<br />

SETUP<br />

TRANSDUCER<br />

PRESELECT<br />

ON OFF<br />

PREAMP<br />

ON OFF<br />

SETUP<br />

FIRMWARE<br />

UPDATE<br />

FSE MODE<br />

ON OFF<br />

The SETUP key opens the menu for<br />

configuration of the ESIB.<br />

The TRANSDUCER softkey opens a<br />

submenu for entering correction characteristics<br />

for the measurement transducer.<br />

Softkey PRESELECT activates the<br />

preselection in the Analyzer mode.<br />

Softkey PREAMP switches the preamplifier on<br />

and off while preselection is active.<br />

The LISN softkey opens the submenu<br />

including the settings required to control V-<br />

networks (LISNs).<br />

The OPTIONS softkey allows the enabling of<br />

firmware options (Application Firmware<br />

Modules).<br />

The REFERENCE INT/EXT and EXTERNAL<br />

REF FREQUENCY softkeys determine the<br />

reference to be used.<br />

The SERVICE softkey provides special setup<br />

functions which, although not necessary in<br />

normal use, are useful in instrument service.<br />

The GENERAL SETUP softkey opens a<br />

submenu for all the general settings such as<br />

date and time as well as the configuration of<br />

the device interfaces.<br />

LISN<br />

OPTIONS<br />

REFERENCE<br />

INT EXT<br />

EXT REF<br />

FREQUENCY<br />

SERVICE<br />

GENERAL<br />

SETUP<br />

The FIRMWARE UPDATE softkey opens a<br />

submenu for the installation of a firmeware<br />

update.<br />

The FSE MODE softkey determines whether<br />

the ESIB is FSE-compatible after preset.<br />

Using Transducers<br />

A transducer is often connected ahead of ESIB both during the measurement of useful signals and <strong>EMI</strong><br />

and converts the useful or interference variable such as field strength, current or RFI voltage into a<br />

voltage across 50 Ohm. Transducers with a frequency-independent transducer factor can be coded in<br />

10-dB steps together with the unit at connector PROBE CODE. They are supplied at the same time by<br />

this connector. Transducers such as antennas, probes or current probes mostly have a frequencydependent<br />

transducer factor which can be stored in ESIB and automatically has the correct unit during<br />

level measurement.<br />

If a transducer is switched on it is considered as part of the unit during the measurement, ie the<br />

measured values are displayed in the correct unit and magnitude. When working with two measurement<br />

windows, the transducer is always assigned to two windows.<br />

ESIB distinguishes between transducer factor and transducer set. A transducer factor takes the<br />

frequency response of a single transfer element, eg an antenna into consideration. A transducer set can<br />

summarize different transducer factors in several subranges (several transducer factors at the same<br />

time), eg an antenna, a cable and a diplexer.<br />

1088.7531.12 4.21 E-15